Abstract

BackgroundIn critically ill patients, clinicians can have difficulty obtaining accurate oximetry measurements. ObjectiveTo compare the accuracy of nasal alar and forehead sensor measurements and incidence of pressure injury. Methods43 patients had forehead and nasal alar sensors applied. Arterial samples were obtained at 0, 24, and 120 hours. Oxygen saturations measured by co-oximetry were compared to sensor values. Skin was assessed every 8 hours. ResultsOxygen saturations ranged from 69.8%-97.8%, with 18% of measures < 90%. Measurements were within 3% of co-oximetry values for 54% of nasal alar compared to 35% of forehead measurements. Measurement failures occurred in 6% for nasal alar and 22% for forehead. Three patients developed a pressure injury with the nasal alar sensor and 13 patients developed a pressure injury with the forehead sensor (χ2 = 7.68; p = .006). ConclusionsIn this group of patients with decreased perfusion, nasal alar sensors provided a potential alternative for continuous monitoring of oxygen saturation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.