Abstract
Five analysis methods were compared for the determination of boron in heavy water moderator: isotope dilution mass spectrometry, spectrophotometry, neutron activation, inductively coupled plasma – atomic emission spectrometry, and ion selective electrode potentiometry. Ten samples were analysed by each method; the results showed close agreement between all of the methods. Only mass spectrometry achieved the required precision (<1% rsd) for samples taken during initial reactor operation, but all of the methods achieved sufficient precision (<10% rsd) for samples taken during normal operation. For samples for which the 10B concentration must be determined, only mass spectrometry and neutron activation are applicable.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.