Abstract

This study compares dual-energy computed tomography (DECT) images of a phantom including different material inserts and with additional lateral titanium or stainless steel inserts, simulating bilateral hip prostheses. Dual-source (DS) and fast kV-switching (FKS) DECT with/without metal artefact reduction (MAR) were compared with regards to virtually monoenergetic CT number accuracy and the depiction of different materials. Streak artefacts were observed between the metal inserts that were more severe with steel compared to titanium inserts. The artefact severity and CT number accuracy depended on the photon energy (keV) for both DECT techniques. While MAR generally increased the CT number accuracy and material depiction within the streak artefacts, it sometimes decreased the accuracy outside the streak artefacts for both DS and FKS. FKS depicted the metal inserts more accurately than DS with regards to both CT numbers and external diameter.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.