Abstract
The study is based on estimate of CTDI and DLP values for patients' dose optimization procedures. Technical parameters were obtained for three groups of randomly selected patients undergoing abdominal CT examinations of 320 patients of age 20-80 years. The measured values were obtained on image data and the standard reference values of various machines were obtained from service manual as part of QC/QA and the recommended values from ICRP publication 103. The mean CTDI and DLP parameters were; 6.33mGy and 936.25mGy respectively. Furthermore, the mean recorded values of CTDIVOL values were well within ICRP recommendation when the protocol was completed in one scan. On the other hand, in the case of multiscan the total CTDIVol was higher than the ICRP recommendations. While the mean DLP values were higher than the recommended value of 780 mGy-cm by ICRP publication 103. Finally, approximately 37% of the total varied CTDI and DLP values were higher than the recommended dose by ICRP publication 103.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Scientific Research in Science and Technology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.