Abstract

Several different Monte Carlo codes are currently being used at proton therapy centers to improve upon dose predictions over standard methods using analytical or semi-empirical dose algorithms. There is a need to better ascertain the differences between proton dose predictions from different available Monte Carlo codes. In this investigation Geant4 and MCNPX, the two most-utilized Monte Carlo codes for proton therapy applications, were used to predict energy deposition distributions in a variety of geometries, comprising simple water phantoms, water phantoms with complex inserts and in a voxelized geometry based on clinical CT data. The Gamma analysis was used to evaluate the differences of the predictions between the codes. The results show that in all the cases the agreement was better than clinical acceptance criteria.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.