Abstract

Characterization of the various chemical aspects of composite polymers is important for quality control of manufactured polymers. In this study, we compared three suitable matrices (α cyano-4-hydroxycinnamic acid [CHCA], 2,5 dihydroxy benzoic acid [2,5-DHB], and dithranol), to characterize various synthetic polymers by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Although the spectra obtained with the CHCA and 2,5-DHB matrices were generally good, in certain samples ghost peaks disappeared only when dithranol was used as the matrix. Furthermore, we examined the use of sodium trifluoroacetate (NaTFA) as an additive to reduce interference by metals and copolymers in the spectra. In conclusion, appropriate selection of a matrix, according to the characteristics of the polymer, and the use of additives to improve sensitivity are important considerations for polymer analysis and development.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.