Abstract

To compare macular thickness measurements and segmentation error rates between Stratus optical coherence tomography (OCT) (Carl Zeiss Meditec, Inc., Dublin, CA), and Fourier-domain OCT (RTVue, Optovue, Inc., Fremont, CA). A retrospective study was performed of 93 normal and pathologic eyes from 79 subjects imaged with both OCT instruments on the same day. Both the macular thickness measurement for each Early Treatment Diabetic Retinopathy Study (ETDRS) zone and the incidence of segmentation error in the central macula between the two instruments were compared. Macular thickness measurements for all nine ETDRS zones were higher with RTVue compared with Stratus OCT (P < .01). Linear regression analysis showed the highest correlation in the central macula (R(2) = 0.88), with progressively lower correlation peripherally. The overall segmentation error rate was 29% with Stratus OCT versus 32% with RTVue (P > .05). Macular thickness measurement was greater with RTVue than with Stratus OCT in all ETDRS areas, with the best correlation seen in the central macula. No difference in segmentation error rate was noted between the two instruments.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.