Abstract

The changes in haemoglobin oxygenation (SO2) occurring in the tuberculin reaction in human skin were measured using macro- and micro-lightguide spectrophotometry and the results compared. A significant difference was found between the measurements from the respective instruments. Demonstrating that the micro-lightguide technique measures only in the most superficial capillaries. Laser Doppler flux (LDF) and transcutaneous oxygen (tcpO2) measurements were also obtained concurrently. At the height of the reaction, heating did not significantly change SO2 or LDF, showing that the vessels in the skin were maximally vasodilated. Although SO2 was increased in the reaction, tcpO2 decreased. This suggests that the infiltrating cells may present a diffusion barrier to oxygen between the capillaries and the tissue cells. This study has shown that micro-lightguide spectrophotometry gives a local picture of intracapillary oxygen supply, which is useful in elucidating the pathophysiological changes occurring during chronic inflammation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.