Abstract

The programming of a CI speech processor involves measurement of electrical threshold and ‘most comfortable loudness’ (MCL) levels on each electrode. However, when the derived ‘map’ is activated the resultant percept is often too loud or too soft. In this situation, adjustments of MCL settings are usually made in order to achieve a comfortable percept.The present study examined the range of volume control adjustments in 24 users of the MED-EL C40/40+ device. Volume control settings varied from 36% to 126% with a mean of 84.3% (SD = 20.4%). Additionally, the output compression function (‘maplaw’) was adjusted to test whether this alternative manipulation could result in a comfortable percept. It was found that this could be achieved satisfactorily, with maplaw values ranging from 10 to 2000. Clinical implications of these findings are discussed. Copyright © 2009 John Wiley & Sons, Ltd.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.