Abstract
The performance of two liquid chromatography-mass spectrometry (LC/MS) interfacing techniques, thermospray (TSP) and atmospheric pressure chemical ionization (APCI), for the analysis of benzo[a]pyrene (BaP) metabolites (hydroxy, epoxy and quinone derivatives) was compared. Interface and detection parameters such as source temperature, eluent composition or flow rate were optimized using negative ion mode. In TSP, the main ions are mostly [M]-, [M−H2O]- or [M+CH3COO]-, whereas APCI gives mainly the [M]- and [M−H]- ions. Quantification was carried out by flow injection. Calibration graphs were linear in the range of 10 ng to 1000 ng in TSP and 0.1 ng to 10 ng in APCI. Detection limits were in the range of 1 ng to 20 ng in TSP and 0.002 ng to 0.2 ng in APCI. The presence of BaP-1,6-dione, BaP-3,6-dione, and BaP-6,12-dione was confirmed in environmental samples of air particulate matter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.