Abstract

Doses for panoramic dental radiography are assessed in terms of the dose–width product (DWP) or dose–area product, which gives a measure of the radiation through a whole exposure. The DWP can be measured using a pencil ionisation chamber (IC) similar to that used for computed tomography dose assessment. However, ICs are sensitive to radiation incident from all directions and so backscatter from the image receptor may increase the recorded dose. This study compares measurements performed using four options: a pencil IC mounted straight on the image receptor, the IC mounted with a steel plate to the rear to standardise scatter conditions, the IC mounted with a steel plate and lead collimators in front to minimise the effect of extra-focal radiation, and a Quart Dido employing a one square centimetre semiconductor detector (SD) designed for panoramic measurements. The results indicate that modification of the current method by incorporating a steel plate reduced the measurement dose by 7% on average, but the reduction was greater for units with semiconductor imaging plates. The measurements with the SD agree more closely with the IC with the steel plate to the rear. An IC with a backing plate to standardise scatter or a suitable SD is recommended for measurement on panoramic dental units.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.