Abstract

ObjectiveAirway management in patients with suspected cervical spine injury is classified as a “difficult airway.” The best device for managing difficult airways is not known. Therefore, we conducted an intubation study simulating patients with cervical spine injury using three devices: a conventional Macintosh laryngoscope, a video laryngoscope (GlideScope), and a fiberoptic bronchoscope (MAF-TM). Success rates, intubation time, and complication rates were compared.MethodsNine physician experts in airway management participated in this study. Cervical immobilization was used to simulate a difficult airway. Each participant performed intubation using airway devices in a randomly chosen order. We measured the time to vocal cord visualization, time to endotracheal tube insertion, and total tracheal intubation time. Success rates and dental injury rates were compared between devices.ResultsTotal tracheal intubation time using the Macintosh laryngoscope, GlideScope, and fiberoptic bronchoscope was 13.3 (range, 11.1 to 20.1), 14.9 (range, 12.7 to 22.3), and 19.4 seconds (range, 14.1 to 32.5), respectively. Total tracheal intubation time differed significantly among the devices (P=0.009). Success rates for the Macintosh laryngoscope, GlideScope, and fiberoptic bronchoscope were 98%, 96%, and 100%, respectively, and dental injury rates were 5%, 19%, and 0%, respectively.ConclusionThe fiberoptic bronchoscope required longer intubation times than the other devices. However, this device had the best success rate with the least incidence of dental injury.