Abstract
Inductively coupled plasma-mass spectrometry (ICP-MS) method EPA 200.8 is gradually finding acceptance as an alternative to uranium analysis. A comparison of the ICP-MS with the accepted radiochemical method EPA 908.0 has been carried out based on data from laboratory control standards, national proficiency test samples, and environmental and drinking water samples from the State of Utah. The method detection limit (MDL) for ICP-MS was determined to be 0.017 microg/L or (0.011 pCi/L), and the minimum reporting limit (MRL) was 0.17 microg/L (MDL x 10) or (0.11 pCi/L). The minimum reporting limit for radiochemical 908.0 method is 1 pCi/L. Our spiked matrix recoveries, spiked blank samples, and reference materials deviate only a few percentage from the listed true values. Results demonstrate that the ICP-MS is a superior analytical tool for the determination of uranium in drinking and environmental waters at concentrations required by the United States Environmental Protection Agency.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.