Abstract

Immunoturbidimetric assays for specific proteins are available on “open system” clinical chemistry analyzers. The analytical performance of nine immunoturbidimetric specific protein assays (C3, C4, CRP, Haptoglobin, IgA, IgG, IgM, RF, and Transferrin) was compared to immunonephelometry. Testing was performed on the Abbott ARCHITECT ci8200 and the Dade Behring BNII nephelometer and evaluated for precision, linearity, limit of detection, prozone phenomenon, method comparison, workflow, and proficiency testing survey comparison. Immunoturbidimetric assays performance was satisfactory for total precision, linearity, limit of detection and the prozone effect was not observed. Method comparison was acceptable for the immunoglobulins, CRP and transferrin but less favorable for the other assays, likely due to methodology and antibody specificity differences. Immunourbidimetric specific protein assays allow for efficient test consolidation on a general purpose clinical chemistry analyzer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.