Abstract

To compare image quality and radiation dose between high-pitch mode (HPM) and low-pitch mode (LPM) CT in young children. Forty-seven children (mean age 35.6 months; range, 0-126 months) underwent 49 CT examinations in HPM or LPM and were divided into high or low respiratory rate (RR) groups. A qualitative image quality was compared between the two modes. The volume CT dose index (CTDIvol) and dose-length product (DLP) were evaluated from the dose reports, and effective doses were assessed using a paediatric phantom. Image quality was generally better for HPM than LPM (diagnostic acceptance score, 4.00 vs. 3.46, P = 0.004); the difference was more prominent in the high RR group (4.00 vs. 3.22, P = 0.001). However, there was no significant difference in the low RR group. The mean DLP value was higher in HPM than LPM (29.48 mGy · cm vs. 23.46 mGy · cm, P = 0.022), while CTDIvol was not significantly different. The total effective radiation dose was 26 % higher in HPM than LPM (1.82 mSv vs. 1.44 mSv). LPM can be considered for paediatric lung evaluation in young children with low RRs to reduce radiation dose while maintaining favourable image quality. • Radiation exposure is higher on high-pitch "Flash spiral mode" than on low-pitch "X-CARE mode". • "Flash spiral mode" generally showed better image quality than "X-CARE mode". • Difference in image quality was more prominent in the high RR group. • There was no difference in image quality in the low RR group. • "X-CARE mode" should be considered in a limited population with low RRs.