Abstract
The objective of this study was to compare the ICRU 38 calculated rectal dose with in vivo dosimetry measured doses in cobalt-60 HDR brachytherapy for cervical cancer. A total of 48 brachytherapy insertions done on 15 patients treated from January to March 2017 at our institution were included in this prospective cross-sectional study. The results demonstrated no significant difference between the computed ICRU rectal point dose and in vivo maximum measured rectal dose ((r) 0.6208, p < 0.0001 [S]; t test p = 0.1578 [NS] 95% CI − 0.78 to 0.46), but a significant difference between ICRU rectal point and in vivo mean measured rectal dose ((r) 0.6033, p < 0.0001 [S]; t test p < 0.0001 [S] 95% CI − 0.81 to 0.35). These findings were seen even when sub-analyzed for the two used fraction sizes of 7 Gy and 8 Gy. The results also showed no significant differences in the maximum ((r) 0.9029, p < 0.0001 [S]; t test p = 0.2576 [NS], 95% CI − 0.21 to 0.06) and mean ((r) 0.9766, p < 0.0001 [S]; t test p = 0.2786 [NS], 95% CI − 0.93 to 0.03) doses taken from treatment planning system assigned dose points coinciding with the imaged probes of the in vivo dosimeter. Overall, this study was able to provide additional evidence that in vivo dosimetry can be validly used in the clinical setting to estimate the dose to the rectum during Co-60 HDR brachytherapy. Use of this technique allows for an additional quality assurance method that can contribute to reductions of errors in dose delivery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.