Abstract

A spectrofluorimetric method and a chromatographic method [high-performance liquid chromatography – ultraviolet (HPLC-UV) detection] were compared for vitamin C analysis of green beans. For HPLC-UV, the determination was performed before [ascorbic acid (AA)] and after reduction with dithiothreitol [AA + dehydroascorbic acid (DHAA)]; for spectrofluorimetry, DHAA was determined after oxidation with activated charcoal. The fluorimetric determination of DHAA before oxidation was also carried out, and showed good precision but poor accuracy (very high recovery percentages). HPLC-UV showed better linearity and sensitivity, while spectrofluorimetry was more precise. Both methods were suitable for green bean samples, and the choice of which one to use depends on the particular interest of the analysis: spectrofluorimetry may be preferable when the total content of vitamin C is required, but not for individual analysis of each form. HPLC-UV is more suitable when the interest of the analysis is the ascorbic acid form as a deterioration index of vegetable products.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.