Abstract

Goat milk standards and cow milk standards were used to calibrate a Fossomatic-300 fluorescent cell counter and Dairylab II infrared milk analyzer for analyses of goat milk samples. Somatic cell counts (SCC) of goat milk were 27% lower (5.5 × 10 5 ml − 1 versus 7.0 × 10 5 ml − 1) when the Fossomatic-300 was calibrated with goat milk SCC standards than with cow milk SCC standards ( P < 0.001). When the Dairylab II was calibrated with goat milk component standards, the levels of fat and protein in milk samples were 0.04% and 0.27% higher, respectively, than with cow milk component standards ( P < 0.001). The data indicated that Fossomatic instruments and infrared milk analyzers must be calibrated with goat milk standards for more reliable and accurate analysis of goat milk.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.