Abstract

✓ This study was conducted to evaluate the geometric distortion of angiographic images created from a commonly used digital x-ray imaging system and the performance of a commercially available distortion-correction computer program. A 12 × 12 × 12—cm wood phantom was constructed. Lead shots, 2 mm in diameter, were attached to the surfaces of the phantom. The phantom was then placed inside the angiographic localizer. Cut films (frontal and lateral analog films) of the phantom were obtained. The films were analyzed using GammaPlan target series 4.12. The same procedure was repeated with a digital x-ray imaging system equipped with a computer program to correct the geometric distortion. The distortion of the two sets of digital images was evaluated using the coordinates of the lead shots from the cut films as references. The coordinates of all lead shots obtained from digital images and corrected by the computer program coincided within 0.5 mm of those obtained from cut films. The average difference is 0.28 mm with a standard deviation of 0.01 mm. On the other hand, the coordinates obtained from digital images with and without correction can differ by as much as 3.4 mm. The average difference is 1.53 mm, with a standard deviation of 0.67 mm. The investigated computer program can reduce the geometric distortion of digital images from a commonly used x-ray imaging system to less than 0.5 mm. Therefore, they are suitable for the localization of arteriovenous malformations and other vascular targets in gamma knife radiosurgery.

Full Text