Abstract

The thermal desorption (TD) technique has potential advantages for monitoring polycyclic aromatic hydrocarbons (PAHs) in the air, e.g., no need for sample preparation. This study aims to compare the performance of two types of TD tubes: the generic Tenax TA tube and a proprietary PAH tube specifically designed for PAH sampling. Full calibrations were performed on a TD-gas chromatography/mass spectrometry (GC/MS) system for 16 priority PAHs. Co-located field samples were collected with two tubes at a flow rate of 200 mL/min for 24 h in eight rounds. In the laboratory, blank Tenax and PAH tubes had trace-level PAHs ranging from 0 to 0.35 ng/tube and 0−0.83 ng/tube, respectively, after 2-hr conditioning at 330 °C. The calibration with Tenax tubes showed an average precision of 7.5%, R2 from 0.9959 to 1.0000, and method detection limits (MDLs) from 0.056 to 3.64 ng/m3. The calibration of PAH tubes showed an average precision of 7.5%, R2 from 0.9945 to 1.0000, and MDLs from 0.065 to 5.91 ng/m3. In the field, blank Tenax and PAH tubes had trace PAH levels of 0.01–0.34 ng/tube and 0.01−0.42 ng/tube, respectively. Both tubes had negligible breakthroughs (<4%) and good precisions (<20%). Co-located Tenax and PAH tubes displayed a good agreement for light PAHs (differences mostly <30%) but high uncertainty for heavy PAHs (difference 35−70%). Both tubes are applicable for PAH monitoring, but the proprietary PAH tube does not outperform the generic Tenax TA tube. More work is still needed to improve the sensitivity of TD methods for heavy PAHs, e.g., selecting appropriate sorbents and adopting advanced instruments.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call