Abstract

The purpose of this study was to compare (a) forces generated during lateral compaction and (b) apical microleakage for nickel-titanium (NiTi) and stainless steel (SS) finger spreaders. Twenty-eight extracted human teeth were instrumented using a standardized rotary instrumentation technique. NiTi and SS #30 spreaders were used to obturate molar roots while the forces generated during obturation were measured on a Universal testing machine. Apical microleakage was determined using a fluid filtration method. There was no significant difference in microleakage between spreaders. NiTi spreaders produced significantly less force than SS spreaders in all specimens (p < 0.001).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.