Abstract

Along with advances in methodological technologies, various assays for detecting high-risk human papillomavirus (HR HPV) have been introduced. The GeneFinder HPV liquid beads microarray PCR kit is one of the recently developed. Our aim was to compare the performance of GeneFinder to Hybrid Capture 2 for detection of HR HPV. A total of 900 cervical swab specimens were obtained. All specimens were submitted for HR HPV detection with Hybrid Capture 2 (HC2) and GeneFinder and then additionally analyzed the discordant or both positive results using restriction fragment mass polymorphism (RFMP) genotyping analysis. Hybrid Capture 2 detected 12.8% cases and GeneFinder detected 15.8% cases with 13 HR HPV types. Also, GeneFinder detected 27.4% cases for 32 detectable HPV types. The overall agreement rate was 93.2% with 0.724 kappa coefficient. Discordant results between these two assays were observed in 56 cases. HC2 showed sensitivity of 83.5% and specificity of 95.9%, while GeneFinder showed sensitivity of 85.4% and specificity of 91.9%. For HPV 16 or HPV 18 detection, GeneFinder showed 95.0% or 66.7% of sensitivity and 99.2% or 100%, respectively. Overall coinfection rate was 15.4% (38/247) in GeneFinder analysis. Considering the high agreement rate with HC2, high sensitivity and the ability to differentiate 32 HPV genotypes including HPV 16/18, GeneFinder could be used as a laboratory testing method for the screening of HPV infections. The use of GeneFinder may also contribute to future research associated with the significance of various HPV types and multiple coinfections.