Abstract

This review article compares PET imaging performance with Gallium-68 (68Ga) and Fluorine-18 (18F). The literature on this topic is scarce; hence in order to complete the published data, Monte Carlo calculations, as well as phantom measurements, were carried out. The qualitative and quantitative differences between 68Ga and 18F imaging were evaluated in terms of spatial resolution, sensitivity, contrast and activity recovery coefficients for both human PET systems and small animal PET scanners. The clinical and pre-clinical implications of these differences are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.