Abstract

The Abbott TDx fluorescence polarization immunoassay (FPIA) system has been evaluated and compared with well-established high performance liquid chromatography (HPLC) for the determination of three anticonvulsant drugs: phenytoin, phenobarbitone and carbamazepine. These assays were evaluated for precision, calibration curve stability, specificity and accuracy. Within-run precision studies using control samples (n = 15) in the subtherapeutic, therapeutic, and toxic concentrations, resulted in coefficients of variation in the range of 1.79-3.99% (FPIA) and 1.16-2.52% (HPLC), respectively. Between-run precision ranged from 2.32-6.34% for FPIA and from 2.04-3.38% for HPLC. Comparison of 122 patient samples assayed with both methods indicated an extremely good analytical correlation (r = 0.96) for all three comparisons. The FPIA method offers significant advantages in calibration curve stability while maintaining accuracy and precision comparable with those of established HPLC procedures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.