Abstract

Five CE-marked immunochromatography assay kits for the rapid detection of norovirus were compared with the reference method of real-time polymerase chain reaction (PCR) for the detection of acute norovirus infection. The sensitivity of the assays ranged between 23% and 59% with specificity being 75‒100% when compared with real-time PCR. The data suggest that a reactive immunochromatographic assay indicates norovirus infection. However, the sensitivity of the assay would need to be significantly improved to be suitable for routine diagnostic purposes or as an alternative to laboratory analysis for near-patient testing.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.