Abstract

The density of neutral oxygen atoms in a plasma postglow chamber was measured with a standard Ni catalytic probe and a fiber optical catalytic probe. A highly dissociated oxygen plasma was created in the discharge chamber with an inductively coupled rf generator with a frequency of 27.12 MHz and output power of 200 W. The measurements were performed in the postglow chamber where a movable recombinator for oxygen atoms was placed. The recombinator enabled the adjustment of O density independently of discharge parameters. The density of neutral oxygen atoms was determined at different pressures and different positions of the recombinator. The O density was between 5×1020 and 2×1021 m−3. The fiber probe expressed excellent repeatability as compared to the standard probe. As long as the temperature of the catalyst was above 100 °C both probes gave similar results. At a lower temperature of the standard probe, however, the optical fiber probe gave more reasonable results. The discrepancy was attributed to the decrease of the catalytic activity of nickel at lower temperatures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.