Abstract

Commercial planning systems used in radiotherapy treatments use determinist correlations to evaluate dose distribution around regions of interest. Estimated dose with this type of planners can be problematic, especially when analyzing heterogeneous zones. The present work is focused in quantifying the dose distribution in a heterogeneous medium irradiated by a 6 MeV photon beam emitted by an Elekta Precise Radiotherapy Unit head. Dose mapping inside the heterogeneous water phantom has been simulated with the photon and electron transport with Monte Carlo computer code MCNP5 and also, using a commercial treatment planning software in the same irradiation conditions. The calculated results were compared with experimental relative dose curves. This comparison shows that inside the heterogeneity region, the commercial algorithms are not able to predict the variation of dose in the heterogeneous zones with the same precision as MCNP5.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.