In total, 172 isolates of Enterobacteriaceae, Acinetobacter spp., Pseudomonas aeruginosa and Stenotrophomonas maltophilia were tested for susceptibility to colistin by agar dilution, Etest and the Vitek 2 system. Isolates with a colistin MIC < or =2 mg/L were considered to be susceptible. Fifty-four (31%) Gram-negative isolates were resistant to colistin. Categorical agreement between agar dilution and Etest was 87%, and between agar dilution and Vitek 2 was 82%. Based on the data obtained, the Vitek 2 system was unreliable for detecting colistin resistance, and results obtained by Etest may require confirmation by a standard MIC susceptibility testing method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.