Abstract

To compare the analytical performances of the enzymatic method (EM) and capillary electrophoresis (CE) for hemoglobin A1c (HbA1c) measurement. Imprecision, carryover, stability, linearity, method comparison, and interferences were evaluated for HbA1c via EM (Abbott Laboratories, Inc) and CE (Sebia). Both methods have shown overall within-laboratory imprecision of less than 3% for International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) units (<2% National Glycohemoglobin Standardization Program [NGSP] units). Carryover effects were within acceptable criteria. The linearity of both methods has proven to be excellent (R2 = 0.999). Significant proportional and constant difference were found for EM, compared with CE, but were not clinically relevant (<5 mmol/mol; NGSP <0.5%). At the clinically relevant HbA1c concentration, stability observed with both methods was acceptable (bias, <3%). Triglyceride levels of 8.11 mmol per L or greater showed to interfere with EM and fetal hemoglobin (HbF) of 10.6% or greater with CE. The enzymatic method proved to be comparable to the CE method in analytical performances; however, certain interferences can influence the measurements of each method.