Abstract

Gamma-ray energy spectra were acquired using a Multichannel Analyzer (MCA) and a digitizer. A spectrometer which was consisted of a NaI(Tl) inorganic scintillation detector and a digitizer were used to obtain energy spectra of 137Cs radioisotope. The energy resolution values from analog and digital ways were calculated and compared with each other. The average analog and digital energy resolution values were determined as 7.69 % and 6.80 %, respectively. Obtained results showed that the digital systems are more preferable than the analog ones from the energy resolution point of view

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.