Abstract

Carbamazepine is a widely used anti-epileptic drug with narrow therapeutic range. Many methods have been developed for monitoring the serum drug level. Differential pulse voltammetry (DPV), an electrochemical method advantaged by simple, inexpensive, and relatively short analysis time, has recently been developed for carbamazepine detection. We used a newly developed DPV method with glassy carbon as a working electrode to determine the carbamazepine level. The performance of DPV is compared with the widely used fluorescence polarization immunoassay (FPIA) technique in precision, accuracy, linearity and detection limit. The precision, linearity and accuracy of the DPV and FPIA techniques were comparable at most clinical used levels. The detection limit was 1 μg/mL for the DPV technique and 0.5 μg/mL for the FPIA technique. The performance of the DPV technique was within the FDA guidelines for bioanalytical methods, which ensures the clinical applicability of the DPV technique. The DPV technique may have the potential to be a good alternative for carbamazepine analysis.