In this work, we compare three types of transceive wireless coils: Helmholtz-type coil, metamaterial-inspired coil and their combination. Each transceive coil is electromagnetically coupled to a body coil “birdcage” type of a 1.5 T MR scanner and improves bilateral breast imaging performance. While Helmholtz-type coil and metamaterial-inspired coil based on coupled split loop resonators are linearly polarized, their combination allows to couple with both linear components of the radiofrequency magnetic field, providing a more significant effect of a local boosting of the body coil’s transmit efficiency and radiofrequency safety in comparison with birdcage coil only.