Abstract

Light transmission aggregometry (LTA), the gold standard for the study of patients with defects of platelet function, is a poorly standardized technique. The guidelines that have been produced so far are largely based on consensus of experts, due to the absence of studies directly comparing different procedures. Therefore, ad hoc studies are needed to gather scientific evidence on how to choose the most appropriate procedures for LTA measurement. In this study, we aimed at evaluating the most appropriate conditions for preparing samples of platelet-rich plasma (PRP) for studies of platelet aggregation by LTA. Citrate-anticoagulated blood from 32 individuals was centrifuged at 150, 200, 250 or 300×g at room temperature for 10 min. Red blood cells contamination was highest in PRP prepared at 150×g; mean platelet volume (MPV) was lowest in PRP prepared at 300×g. The extent of platelet aggregation measured by LTA was lower and more variable in PRP prepared at 300×g. Therefore, centrifugation of blood at 200×g or 250×g for 10 min appears to be the best condition for preparing PRP for LTA studies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.