Abstract

A study was carried out in which 50 μm I.D. fused-silica capillaries were packed with 3 μm octadecylsilane bonded silica, from the same batch, by four methods; liquid slurry and carbon dioxide supercritical carrier, each with and without the use of an ultrasonic probe. A neutral test mixture was analysed by capillary column in reversed-phase mode, and the reproducibility of the electroosmotic flow and of migration time, column efficiency and retention factors, was determined. Initially results suggested that there was no significant difference between properties of columns packed by different methods, and a more thorough statistical evaluation confirmed this; differences observed in the column performance were attributed to random variations between replicate columns, and not between packing methods. However, the variation was least when applying the ultrasonication during liquid slurry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.