Abstract
A comparative study of different chemical modifiers for the direct determination of arsenic in real and in synthetic sea water samples of high (72.8 per thousand ) and low (34.2 per thousand ) salinity was carried out. The use of lanthanum chloride and magnesium nitrate (LOD=3.0 microg l(-1)) offers good recoveries for low salinities, while for high salinities an enhancement higher than 200% was obtained, rendering this method unsuitable for sea water samples. The use of silver nitrate (LOD=2.0 microg l(-1)) produces unfavourable analytical recoveries, around 150 and 300% for synthetic sea water of low and high salinity. Zirconium oxychloride (LOD=1.8 microg l(-1)) and palladium nitrate alone or combined with magnesium nitrate and reducing agents (LOD=1.1-1.3 microg l(-1)) produce analytical recoveries close to 100% for all arsenic concentration tried over all salinities; thus, a single calibration curve using aqueous standard solutions may be applied to the analysis of sea water samples over all salinities. Finally, the addition of reducing agents and magnesium nitrate to palladium does not improve the sensitivity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.