Abstract
X-ray spectra were measured using a CdZnTe detector for the various combinations of water phantom thickness (15–25 cm), iodine contrast media thickness (3–60 μm in a 1 cm thick simulated tumour), La filter thickness (0, 100, and 160 μm), and X-ray tube voltage (50, 65, and 80 kV). The iodine thickness was obtained as a contrast using energy information from the X-ray spectra and current, which was calculated as the summed products of X-ray energy and events. The most sensitive method of iodine detection is the energy measurement method with 65 kV tube voltage.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.