Abstract

In clinical tests, the degree of 226 stenoses in 200 patients undergoing prima vista PTCA was evaluated digital on-site and on-line by both, automated/manual (EDa/EDm) edge detection method and crossline stenosis method (CL). The analysis was performed by use of AWOS V4.01, an angiographic workstation for digital quantitative coronary angiography. A tight correlation between EDa and EDm was observed (r=0.99, p/spl les/0.0001). In addition, excellent correlations between diameter measurements by CL and both EDa and EDm were observed (for both: r=0.97, p/spl les/.0001). It is concluded that both, automated/manual edge detection method and crossline stenosis method compare favorably for measuring the minimal lumen diameter enabling onsite as well as decision making for interventional therapy as reliable evaluation of the morphological results after such procedures in a digital catheterization laboratory. >

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.