Abstract

To compare corneal astigmatism measurements obtained using a new optical biometer (AL-Scan) and a standard biometer based on partial coherence interferometry (IOLMaster 500) for toric intraocular lens (IOL) selection. Siriraj Hospital, Mahidol University, Bangkok, Thailand. Prospective comparative study. Keratometry (K) readings were taken with the 2 biometers in eyes of cataract patients. The corneal astigmatism readings from the 2 devices were compared. The cylinder measurements were transformed into J0 and J45 vector components for analysis, and the devices' values were compared using the paired t test. Correlation coefficients were calculated. The Bland-Altman method was used to evaluate the agreement between the sets of measurements. One hundred thirty-seven eyes (81 patients) were enrolled. The magnitude of astigmatism between the new biometer 2.4 mm zone and the standard biometer was similar (P = .19). There was a statistically significant difference between values from the new biometer 3.3 mm zone and the standard biometer (P < .05). There was good correlation in the overall magnitude of astigmatism between devices. There was a trend of low correlation toward the flat cornea (K <43.0 diopters [D]), low corneal cylinder (<1.0 D), and with-the-rule astigmatism. The IOL toricity calculated from both devices was similar. Corneal astigmatism measurements with the 2 biometers were similar in magnitude, especially in 2.4 mm zone of the new biometer. There was no difference in the toric IOL selections between the devices. No author has a financial or proprietary interest in any material or method mentioned.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call