Abstract

Multiple factors have been identified as contributing to polyethylene wear and debris generation of the acetabular lining. Polyethylene wear is the primary limiting factor in the functional behavior and consequent longevity of a total hip arthroplasty (THA). This retrospective study reviewed the clinical and radiographic data of 77 consecutive THAs comparing in vivo polyethylene wear of two similar acetabular cup liners. Minimum follow-up was 7 years (range 7–15). The incidence of measurable wear in a group of machined liners sterilized with ethylene oxide and composed of GUR 1050 stock resin was significantly higher (61%) than the compression-molded, GUR 1020, O2-free gamma irradiation sterilized group (24%) (P = 0.0004). Clinically, at a 9-year average followup, both groups had comparable HHS scores and incidence of thigh or groin pain, though the machined group had an increased incidence of osteolysis and annual linear wear rate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.