Abstract

To compare the continuous real time blood glucose (CG) measurement with venous laboratory blood glucose (LG) level in neonates during perioperative period. Glucose levels were measured simultaneously by CG, glucometer glucose (GG) and LG at 40 time points in ten neonates during perioperative period. Intraclass correlation coefficient (ICC) and Bland Altman analysis were used for comparison. Correlation between CG and LG was excellent (ICC= 0.953; P<0.001), and average difference was 23.8 (95%CI 52.9 to -5.3) mg/dL, showed better reliability than at hyperglycemic state (ICC=0.653; P=0.006). The GG-LG showed excellent reliability with ICC = 0.985; P<0.001 and average difference of 15.4 (95% CI 30.7 to 0.1) mg/dL. CG at euglycemic state (ICC= 0.880; P<0.001). CG measurement is reliable for blood sugar estimation in neonates; but has lower reliability for hyperglycemia. The continuous trend of glucose measurement by CG is helpful for timely diagnosis of hyperglycemia during perioperative period in neonates.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.