Abstract

A wearable electrocardiogram(ECG) device using conductive fabric sensor was compared with traditional Ag-AgCl electrode ECG device. The ECG signals were measured under existence of motion artifacts on variable running speed using treadmill to verify that wearable device can substitute traditional ECG device. A signal to noise ratio (SNR) and RR interval were compared between the two devices. The SNR of wearable device was similar or higher than that of clinical device and difference of RR interval was 2ms. The results show that the wearable ECG device using conductive fabric sensor can make similar performance with ECG device using Ag-AgCl electrode even under motion artifacts.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.