Abstract

Braided stents are associated with a number of complications in vivo. Accurate computational modelling of these devices is essential for the design and development of the next generation of these stents. In this study, two commonly utilised methods of computationally modelling filament interaction in braided stents are investigated: the join method and the weave method. Three different braided stent designs are experimentally tested and computationally modelled in both radial and v-block configurations. The results of the study indicate that while both methods are capable of capturing braided stent performance to some degree, the weave method is much more robust.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.