Abstract
The addition of junctions in modular stems implies a greater susceptibility to corrosion. The aim of this study is to compare serum chromium and cobalt levels after using a bimodular stem and its monoblock counterpart in primary total hip arthroplasty. Postoperative clinical scores were also compared. A prospective cohort study between 2012 and 2015 was designed. One arm of the cohort included patients with the cementless modular neck stem H-Max M® and the other with the cementless monoblock stem counterpart H-Max S®. No statistically significant difference was found in chromium value between groups (p = 0.621) at two years postoperative. Cobalt value was higher in the modular group (p = < 0.001). No statistically significant difference was found in clinical postoperative scores except for the Harris Hip Score, with better results at six months in modular group (p = 0.007). Higher serum cobalt level in the modular group has limited the use of modular stems in our daily practice. Advantages of modular stem were not found. II.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: European journal of orthopaedic surgery & traumatology : orthopedie traumatologie
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.