Abstract

Commissioning data of the common energy of 6 MV photon beams are analyzed for seven Philips/ Elekta linacs of five different models: SL 75/5, SL 15, SL 25, SLi Precise, and SLi. The observed beam quality %dd(10)x of 67.7% +/- 0.3 agrees with the manufacturer's specification of 67.5% +/- 2% but with a much smaller variation, suggesting the possibility of a more accurate beam quality guideline for acceptance commissioning of these linac models. Field size factors are less uniform, especially for the biggest field size of 40 x 40 cm2, but all agree within 1.1% between the five models. These 6 MV photon beams are also shown to be 0.4 MV harder than the Varian 2100C model. For commissioning electron beams generated by the Philips/Elekta models SL 15, SLi Precise and SLi, guidelines can be found from this work for the following parameters: actual field size at 100 SSD (up to +/- 0.6%) (different from manufacturer specified values at 95 SSD), and virtual source distance (up to +/- 0.7%). Significant differences are seen for R50 (up to 3.7 mm), sigma-theta-x, effective source distance, and cone size factor. However, except for R50 where Varian 2100C datasets are not available for comparison, the differences between the different Philips/Elekta models are comparable to those between the same Varian 2100C model linacs measured by us and Watts.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.