Abstract

We compared in vitro rates of hemolysis for a recently developed centrifugal pump with a conventional roller pump (10-10-00; Stöckert, Munich, Germany). Flow rates of 0.3 L/min and 1 L/min and a pressure of 200 mm Hg were chosen to simulate conditions during neonatal extracorporeal membrane oxygenation (ECMO). There was no significant difference in hemolysis rates between centrifugal and roller pumps (p = 0.57) nor between high and low flow (p = 0.86). The centrifugal pump caused no more blood trauma than the roller pump at the low-flow/high-pressure conditions required for neonatal ECMO. The Nikkiso pump is superior to roller pumps in size and priming volume (25 ml) and may permit development of a smaller and simpler ECMO system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.