Abstract

A few studies have shown that cementing the stem enhances fixation of the tibial baseplate in total knee replacement (TKR). Even the horizontal technique has been shown to provide good fixation. We used radiostereometry to study migration of the tibial component in 30 knees operated with Profix TKR. The knees were randomised for either complete (both under the baseplate and around the stem) or horizontal (only under the baseplate) cementing of the tibial component. At two years the tibial baseplate rotated externally a median of 0.18 degrees in the uncemented stem group and internally a median of 0.23 degrees in the cemented stem group. The tibial baseplate subsided 0.14 mm in the cemented stem group, and no translation was seen in the uncemented stem group. The differences in migration were small and probably without clinical significance. The findings do not favour either of the cementing techniques in TKR.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call