Abstract

Thrombogenicity testing is a key component in the development of medical devices intended for contact with blood. The Chandler loop system has previously been used as an in vitro thrombogenicity testing method. In this study, we used a modified version of the Chandler loop model to evaluate its capacity to detect differential thrombogenic potential of different catheter materials using goat blood. We also sought to determine the optimal experimental conditions for detecting the thrombogenicity of catheter material. Using the Chandler loop system with goat blood we demonstrated that silicone catheters had a stronger thrombogenicity as compared to polyurethane catheters as evidenced by significantly larger thrombi (p < 0.001) and higher infusion pressures (p < 0.05). This is consistent with many, but not all, previous in vitro and in vivo studies comparing polyurethane to silicone catheters. The use of this modified Chandler loop system with goat blood may provide an additional in vitro testing platform for thrombogenicity testing of catheters. © 2018 Wiley Periodicals, Inc. J Biomed Mater Res Part A: 106A: 3143-3151, 2018.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.