Abstract

Measuring seroprevalence over time is a valuable epidemiological tool for improving our understanding of COVID-19 immunity. Due to the large number of collections required for population surveillance as well as concerns about potential infection risk to the collectors, self-collection approaches are being increasingly pursued. To advance this methodology, we collected paired venous and capillary blood samples by routine phlebotomy and Tasso-SST device respectively from 26 participants and measured total immunoglobulin (Ig) and IgG antibodies to the SARS-CoV-2 receptor binding domain (RBD) by enzyme-linked immunosorbent assay (ELISA) on both specimens. Qualitatively, no discrepancies were noted in binary results between Tasso and venipuncture-derived plasma. Furthermore, in vaccinated participants, correlation between Tasso and venous total Ig and IgG specific antibody quantitative levels was high (Total Ig: Spearman ρ = 0.72, 95% CI (0.39,0.90); IgG: Spearman ρ = 0.85, 95% CI (0.54, 0.96)). Our results support the use of Tasso at-home collection devices for antibody testing.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call