Abstract

For the measurement of soil bulk electrical conductivity using time domain reflectometry (TDR), it is necessary to determine cable and probe properties. This can be done by direct measurement or by calibration of the TDR system in salt solutions. For a three‐wire probe system, the results showed that with least‐squares fitting, the electrical conductivity of salt solutions could be measured most accurately. Comparison of several calibration techniques suggested that directly measured cable properties deviate from their calibrated (optimal) values. The difference in accuracy between calibration and direct measurements of cable properties could not be explained with overfitting, as has been suggested. It was concluded that a reduction of calibration time by direct measurement of probe and cable properties is not advisable in case of three‐wire probes. When necessary, a reduction of calibration time can be achieved by limiting the calibration procedure to two well chosen combinations of cable length and solution concentration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.