Abstract

A comparison between C(18) silica and multi-walled carbon nanotubes (MWCNTs) in the extraction of Chlorpyrifos and Phosalone in environmental water samples was carried out using HPLC. Parameters affecting the extraction were type and volume of elution solvent, pH and flow rate of sample through the adsorbent. The optimum conditions obtained by C(18) cartridge for adsorption of these pesticides were 4 mL dichloromethane as elution solvent, sample pH of 5, flow rate of 1 mL/min, and those for MWCNT cartridge were 3 mL dichloromethane, pH of 5 and flow rate of 10 mL/min, respectively. Optimized mobile phase for separation and determination of these compounds by HPLC was methanol/water (80:20 v/v) with pH=5 (adjusted with phosphate buffer). Under optimal chromatographic and SPE conditions, LOD, linear range and precision (RSD n=8) were 3.03x10(-3), 0.01-5.00 microg/mL and 2.7% for Chlorpyrifos and 4.03x10(-4), 0.01-5.00 microg/mL and 2.3% for Phosalone, in C(18) cartridge, respectively. These values for MWCNT were 4.02x10(-6), 0.001-0.500 microg/mL and 1.8% for Chlorpyrifos and 1.02x10(-6), 0.001-0.500 microg/mL and 1.5% for Phosalone, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.